o2_sensor
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| o2_sensor [2025/01/17 02:43] – kenson | o2_sensor [2025/01/30 20:43] (current) – kenson | ||
|---|---|---|---|
| Line 3: | Line 3: | ||
| * {{ :ocr60.pdf | Whitepaper describing use of 9, | * {{ :ocr60.pdf | Whitepaper describing use of 9, | ||
| * The oxygen quenching is predicted by the Stern-Volmer equation, and changes in intensity or lifetime of the fluorescence can be monitored | * The oxygen quenching is predicted by the Stern-Volmer equation, and changes in intensity or lifetime of the fluorescence can be monitored | ||
| + | * Ruthenium complex | ||
| + | * [[https:// | ||
| + | * [[https:// | ||
| + | * [[https:// | ||
| + | * {{: | ||
| + | * [[https:// | ||
| + | * [[https:// | ||
| + | * Datasheet: [[https:// | ||
| + | * [[https:// | ||
| + | * {{ : | ||
| + | [Ru(bipy)3]2+, | ||
| ==== 390nm LED ==== | ==== 390nm LED ==== | ||
| Line 49: | Line 60: | ||
| {{: | {{: | ||
| - | ==== Making | + | ===== Making |
| + | |||
| + | ==== Doping rubrene into PDMS ==== | ||
| + | (e.g., Sylgard 184 or a similar PDMS variant like Sylcap 284). | ||
| - | Doping rubrene into PDMS (e.g., Sylgard 184 or a similar PDMS variant like Sylcap 284) | ||
| Most published oxygen-sensor designs with rubrene start in a relatively narrow doping range to balance brightness and avoid dye aggregation or self-quenching. | Most published oxygen-sensor designs with rubrene start in a relatively narrow doping range to balance brightness and avoid dye aggregation or self-quenching. | ||
| - | ## 1. Typical Doping Range ## | + | ==== 1. Typical Doping Range ==== |
| * **Starting Point**: **0.01–0.1 wt%** (by weight of rubrene relative to the total weight of uncured PDMS) is a commonly cited window. | * **Starting Point**: **0.01–0.1 wt%** (by weight of rubrene relative to the total weight of uncured PDMS) is a commonly cited window. | ||
| * In practical terms, this might translate to about **0.1–1 mg of rubrene per gram of PDMS base**. | * In practical terms, this might translate to about **0.1–1 mg of rubrene per gram of PDMS base**. | ||
| * If you see **aggregation** or **self-quenching** (the fluorescence drops because the molecules are packed too closely), reduce the concentration. If the **signal is too weak**, you can increase it slightly toward the upper end of the range. | * If you see **aggregation** or **self-quenching** (the fluorescence drops because the molecules are packed too closely), reduce the concentration. If the **signal is too weak**, you can increase it slightly toward the upper end of the range. | ||
| - | ### Why This Range? | + | === Why This Range? |
| - | - **Fluorescence Brightness vs. Self-Quenching** | + | - **Fluorescence Brightness vs. Self-Quenching** |
| * Rubrene is highly fluorescent, | * Rubrene is highly fluorescent, | ||
| * Above roughly **0.1–0.2 wt%**, many organic dyes (including rubrene) begin to exhibit “concentration quenching, | * Above roughly **0.1–0.2 wt%**, many organic dyes (including rubrene) begin to exhibit “concentration quenching, | ||
| - **Cost and Solubility** | - **Cost and Solubility** | ||
| - | | + | * Rubrene can be expensive, and it’s only moderately soluble in typical organic solvents. Over-saturating the mix will result in precipitates, |
| - | + | ||
| - | 3. **Oxygen Diffusion** | + | |
| - | - PDMS is highly oxygen-permeable, | + | |
| - | + | ||
| - | --- | + | |
| - | + | ||
| - | ## 2. Practical Mixing Tips | + | |
| - | 1. **Pre-Dissolve the Rubrene** | + | - **Oxygen Diffusion** |
| - | | + | * PDMS is highly oxygen-permeable, which is great for oxygen sensing but also means the dye gets quenched quickly. Going too high in dye content doesn’t necessarily improve sensitivity—it may just cause more self-quenching without improving |
| - | - Slowly add this solution to the PDMS base resin while stirring. This helps achieve more uniform dispersal. | + | |
| - | 2. **Degas** | + | ==== 2. Practical Mixing Tips ==== |
| - | - After mixing, place the solution under vacuum to remove both solvent and air bubbles. | + | |
| - | - If you have a two-part system (base + curing agent), you might do a preliminary degassing before adding the curing agent, then a final degassing after thoroughly mixing in the catalyst. | + | |
| - | 3. **Curing** | + | - **Pre-Dissolve the Rubrene** |
| - | - Many PDMS systems can cure at room temperature | + | * Dissolve rubrene |
| - | - Try not to exceed 120 °C or keep the mixture at high temperature for too long—rubrene will degrade over time if both heat and oxygen are present. | + | * Slowly add this solution |
| - | --- | + | |
| + | * After mixing, place the solution under vacuum to remove both solvent and air bubbles. | ||
| + | * If you have a two-part system (base + curing agent), you might do a preliminary degassing before adding the curing agent, then a final degassing after thoroughly mixing in the catalyst. | ||
| - | ## 3. Considerations | + | - **Curing** |
| + | * Many PDMS systems can cure at room temperature in 24–48 hours or at 60–70 °C for a few hours. | ||
| + | * Try not to exceed 120 °C or keep the mixture at high temperature for too long—rubrene will degrade over time if both heat and oxygen are present. | ||
| - | 1. **Absorption Spectrum of Rubrene** | + | ==== 3. Considerations for a 450 nm LED ==== |
| - | - Rubrene has a primary absorption peak around ~495–530 nm. However, it still absorbs sufficiently at ~450 nm to fluoresce orange. You won’t be hitting its absolute peak absorption, but in practice, 450 nm can still excite rubrene decently. | + | |
| - | - You may consider verifying the absorbance in your final PDMS film to ensure you’re not “under-absorbing.” | + | |
| - | 2. **Film Thickness** | + | - **Absorption Spectrum of Rubrene** |
| - | - If the film is too thick or dye concentration is too high, inner regions might not be excited properly due to light attenuation. | + | * Rubrene absorbs sufficiently at ~450 nm to fluoresce orange. |
| - | - A thickness of **~0.1–1 mm** is typical for sensor films, but this depends on optical design and the LED intensity. | + | * Verifying |
| - | 3. **Sensor Calibration** | + | - **Film Thickness** |
| - | - Once cured, measure | + | * If the film is too thick or dye concentration is too high, inner regions might not be excited properly due to light attenuation. |
| + | * A thickness of **~0.1–1 mm** is typical for sensor films, but this depends on optical design and the LED intensity. | ||
| - | --- | + | |
| + | * Once cured, measure the fluorescence intensity (and/or lifetime) under different known oxygen concentrations to build a calibration curve using the Stern–Volmer relationship. | ||
| - | ## 4. Adjusting the Ratio Over Time | ||
| - | - **If the Film Is Too Dark (Self-Absorbing)** | + | ==== 4. Adjusting |
| - | - Lower the dye percentage. Even going from 0.1 wt% down to 0.02 wt% can make a big difference in clarity and reduce self-quenching. | + | |
| - | - **If the Signal Is Too Dim** | + | |
| - | - Increase the dye content in small increments (e.g., from 0.02 wt% up to 0.05 wt%), but watch for diminishing returns due to concentration quenching. | + | |
| - | --- | + | * **If the Film Is Too Dark (Self-Absorbing)** |
| + | * Lower the dye percentage. Even going from 0.1 wt% down to 0.02 wt% can make a big difference in clarity and reduce self-quenching. | ||
| + | * **If the Signal Is Too Dim** | ||
| + | * Increase the dye content in small increments (e.g., from 0.02 wt% up to 0.05 wt%), but watch for diminishing returns due to concentration quenching. | ||
| - | ### Bottom Line | + | ==== Bottom Line ==== |
| A good **starting point** is **0.01–0.1 wt% rubrene** relative to the PDMS base. Begin on the lower side (e.g., **0.02–0.05 wt%**) to avoid aggregation, | A good **starting point** is **0.01–0.1 wt% rubrene** relative to the PDMS base. Begin on the lower side (e.g., **0.02–0.05 wt%**) to avoid aggregation, | ||
| - | ==== pH ==== | + | ==== pH optode sensor notes ==== |
| * {{ : | * {{ : | ||
| * Pyranine Solvent Green 7 CAS 6358-69-6 aka HPTS aka Pyranine | * Pyranine Solvent Green 7 CAS 6358-69-6 aka HPTS aka Pyranine | ||
o2_sensor.1737081828.txt.gz · Last modified: 2025/01/17 02:43 by kenson
